Abstract

This NeuroLibre Reproducible preprint is an interactive tutorial on quantitative T1 mapping MRI. It is an interactive version of two subsections of the chapter “Quantitative T1 and T1r Mapping” in the book Quantitative Magnetic Resonance Imaging Boudreau et al., 2020.

\awesomebox[red]{2pt}{\faExclamationCircle}{red}{\textbf{NOTE}}

The following section in this document repeats the narrative content exactly as found in the corresponding NeuroLibre Reproducible Preprint (NRP). The content was automatically incorporated into this PDF using the NeuroLibre publication workflow Karakuzu et al., 2022 to credit the referenced resources. The submitting author of the preprint has verified and approved the inclusion of this section through a GitHub pull request made to the source repository from which this document was built. Please note that the figures and tables have been excluded from this (static) document. To interactively explore such outputs and re-generate them, please visit the corresponding NRP. For more information on integrated research objects (e.g., NRPs) that bundle narrative and executable content for reproducible and transparent publications, please refer to DuPre et al. (2022). NeuroLibre is sponsored by the Canadian Open Neuroscience Platform (CONP) Harding et al., 2023.

This Jupyter Book is a series of interactive tutorials about quantitative T1 mapping, powered by qMRLab Karakuzu et al., 2020.

Widely considered the gold standard for T1 mapping, the inversion recovery technique estimates T1 values by fitting the signal recovery curve acquired at different delays after an inversion pulse (180°). In a typical inversion recovery experiment (Figure 1), the magnetization at thermal equilibrium is inverted using a 180° RF pulse. After the longitudinal magnetization recovers through spin-lattice relaxation for predetermined delay (“inversion time”, TI), a 90° excitation pulse is applied, followed by a readout imaging sequence (typically a spin-echo or gradient-echo readout) to create a snapshot of the longitudinal magnetization state at that TI.

Inversion recovery was first developed for NMR in the 1940s Hahn, 1949Drain, 1949, and the first T1 map was acquired using a saturation-recovery technique (90° as a preparation pulse instead of 180°) by Pykett & Mansfield, 1978. Some distinct advantages of inversion recovery are its large dynamic range of signal change and an insensitivity to pulse sequence parameter imperfections Stikov et al., 2015. Despite its proven robustness at measuring T1, inversion recovery is scarcely used in practice, because conventional implementations require repetition times (TRs) on the order of 2 to 5 T1 Steen et al., 1994, making it challenging to acquire whole-organ T1 maps in a clinically feasible time. Nonetheless, it is continuously used as a reference measurement during the development of new techniques, or when comparing different T1 mapping techniques, and several variations of the inversion recovery technique have been developed, making it practical for some applications Messroghli et al., 2004Piechnik et al., 2010.

M_z(TI)=M_01−cos(θ_180)e−TRT_1−[1−cos(θ_180)]e−TIT_11−cos(θ_180)cos(θ_90)e−TRT_1M\_{z}(TI) = M\_0 \frac{1-\text{cos}(\theta\_{180})e^{- \frac{TR}{T\_1}} -[1-\text{cos}(\theta\_{180})]e^{- \frac{TI}{T\_1}}}{1 - \text{cos}(\theta\_{180}) \text{cos}(\theta\_{90}) e^{- \frac{TR}{T\_1}}}

where Mz is the longitudinal magnetization prior to the θ90 pulse. If the in-phase real signal is desired, it can be calculated by multiplying Eq. 1 by ksin(θ90)e-TE/T2, where k is a constant. This general equation can be simplified by grouping together the constants for each measurements regardless of their values (i.e. at each TI, same TE and θ90 are used) and assuming an ideal inversion pulse:

M_z(TI)=C(1−2e−TIT_1+e−TRT_1)M\_z(TI) = C(1-2e^{- \frac{TI}{T\_1}} + e^{- \frac{TR}{T\_1}})
M_z(TI)=C(1−2e−TIT_1)M\_z(TI) = C(1-2e^{- \frac{TI}{T\_1}})
S(TI)=a+be−TIT_1S(TI) = a + be^{- \frac{TI}{T\_1}}

where a and b are complex values. If magnitude-only data is available, a 3-parameter model can be sufficient by taking the absolute value of Eq. 4. While the RD-NLS algorithms are too complex to be presented here (the reader is referred to the paper, (Barral et al. 2010)), the code for these algorithms was released open-source along with the original publication, and is also available as a qMRLab T1 mapping model. One important thing to note about Eq. 4 is that it is general – no assumption is made about TR – and is thus as robust as Eq. 1 as long as all pulse sequence parameters other than TI are kept constant between each measurement. Figure 4 compares simulated data (Eq. 1) using a range of TRs (1.5T1 to 5T1) fitted using either RD-NLS & Eq. 4 or a Levenberg-Marquardt fit of Eq. 2.

Figure 5 displays an example brain dataset from an inversion recovery experiment, along with the T1 map fitted using the RD-NLS technique.

Despite a widely acknowledged robustness for measuring accurate T1 maps, inversion recovery is not often used in studies. An important drawback of this technique is the need for long TR values, generally on the order of a few T1 for general models (e.g. Eqs. 1 and 4), and up to 5T1 for long TR approximated models (Eq. 3). It takes about to 10-25 minutes to acquire a single-slice T1 map using the inversion recovery technique, as only one TI is acquired per TR (2-5 s) and conventional cartesian gradient readout imaging acquires one phase encode line per excitation (for a total of ~100-200 phase encode lines). The long acquisition time makes it challenging to acquire whole-organ T1 maps in clinically feasible protocol times. Nonetheless, it is useful as a reference measurement for comparisons against other T1 mapping methods, or to acquire a single-slice T1 map of a tissue to get T1 estimates for optimization of other pulse sequences.

0.1Other Saturation-Recovery T1 Mapping techniques

1Variable Flip Angle T1 Mapping

Variable flip angle (VFA) T1 mapping Christensen et al., 1974Fram et al., 1987Gupta, 1977, also known as Driven Equilibrium Single Pulse Observation of T1 (DESPOT1) Homer & Beevers, 1985Deoni et al., 2003, is a rapid quantitative T1 measurement technique that is widely used to acquire 3D T1 maps (e.g. whole-brain) in a clinically feasible time. VFA estimates T1 values by acquiring multiple spoiled gradient echo acquisitions, each with different excitation flip angles (θn for n = 1, 2, .., N and θi ≠ θj). The steady-state signal of this pulse sequence (Figure 1) uses very short TRs (on the order of magnitude of 10 ms) and is very sensitive to T1 for a wide range of flip angles.

Figure 1. Simplified pulse sequence diagram of a variable flip angle (VFA) pulse sequence with a gradient echo readout. TR: repetition time, θn: excitation flip angle for the nth measurement, IMG: image acquisition (k-space readout), SPOIL: spoiler gradient.

M_z(θ_n)=M_01−e−TRT_11−cos(θ_n)e−TRT_1sin(θ_n)(1)\tag{1} M\_{z}(\theta\_n) = M\_0 \frac{1-e^{- \frac{TR}{T\_1}}}{1-\text{cos}(\theta\_n) e^{- \frac{TR}{T\_1}}} \text{sin}(\theta\_n)

where Mz is the longitudinal magnetization, M0 is the magnetization at thermal equilibrium, TR is the pulse sequence repetition time (Figure 1), and θn is the excitation flip angle. The Mz curves of different T1 values for a range of θn and TR values are shown in Figure 2.

From Figure 2, it is clearly seen that the flip angle at which the steady-state signal is maximized is dependent on the T1 and TR values. This flip angle is a well known quantity, called the Ernst angle Ernst & Anderson, 1966, which can be solved analytically from Equation 1 using properties of calculus:

θ_Ernst=acos(e−TRT_1)(2)\tag{2} \theta\_{Ernst} = \text{acos}(e^{- \frac{TR}{T\_1}})

Figure 4. Signal curves estimated using Bloch simulations for three categories of signal spoiling: (1) ideal spoiling (blue), gradient & RF Spoiling (orange), and no spoiling (green). Simulations details: TR = 25 ms, T1 = 900 ms, Te = 100 ms, TE = 5 ms, 100 spins. For the ideal spoiling case, the transverse magnetization is set to zero at the end of each TR. For the gradient & RF spoiling case, each spin is rotated by different increments of phase (2𝜋 / # of spins) to simulate complete decoherence from gradient spoiling, and the RF phase of the excitation pulse is ɸn = ɸn-1 + nɸ0 = ½ ɸ0(n2 + n + 2) Bernstein et al., 2004 with ɸ0 = 117° Zur et al., 1991 after each TR.

S_nsin(θ_n)=e−TRT_1S_ntan(θ_n)+C(1−e−TRT_1)(3)\tag{3} \frac{S\_n}{ \text{sin}(\theta\_n)} = e^{- \frac{TR}{T\_1}} \frac{S\_n}{ \text{tan}(\theta\_n)} + C (1-e^{- \frac{TR}{T\_1}})

As the third term does not change between measurements (it is constant for each θn), it can be grouped into the constant for a simpler representation:

S_nsin(θ_n)=e−TRT_1S_ntan(θ_n)+C(4)\tag{4} \frac{S\_n}{ \text{sin}(\theta\_n)} = e^{- \frac{TR}{T\_1}} \frac{S\_n}{ \text{tan}(\theta\_n)} + C

With this rearranged form of Equation 1, T1 can be simply estimated from the slope of a linear regression calculated from Sn/sin(θn) and Sn/tan(θn) values:

T_1=−TRln(slope)(5)\tag{5} T\_1 = - \frac{TR}{ \text{ln}(slope)}
θ_corrected=B_1θ_nominal(6)\tag{6} \theta\_{corrected} = B\_1 \theta\_{nominal}

B1 in this context is normalized, meaning that it is unitless and has a value of 1 in voxels where the RF field has the expected amplitude (i.e. where the nominal flip angle is the actual flip angle). Figure 6 displays fitted VFA T1 values from a Monte Carlo dataset simulated using biased flip angle values, and fitted without/with B1 correction.

Figure 6. Mean and standard deviations of fitted VFA T1 values for a set of Monte Carlo simulations (SNR = 100, N = 1000), simulated using a wide range of biased flip angles and fitted without (blue) or with (red) B1 correction. Simulation parameters: TR = 25 ms, T1 = 900 ms, θnominal = 6° and 32° (optimized values for this TR/T1 combination). Notice how even after B1 correction, fitted T1 values at B1 values far from the nominal case (B1 = 1) exhibit larger variance, as the actual flip angles of the simulated signal deviate from the optimal values for this TR/T1 (Deoni et al. 2003).

Figure 7 displays an example VFA dataset and a B1 map in a healthy brain, along with the T1 map estimated using a linear fit (Equations 4 and 5).

Figure 7. Example variable flip angle dataset and B1 map of a healthy adult brain (left). The relevant VFA protocol parameters used were: TR = 15 ms, θnominal = 3° and 20°. The T1 map (right) was fitted using a linear regression (Equations 4 and 5).

It has been well reported in recent years that the accuracy of VFA T1 estimates is very sensitive to pulse sequence implementations Baudrexel et al., 2017Lutti & Weiskopf, 2013Stikov et al., 2015, and as such is less robust than the gold standard inversion recovery technique. In particular, the signal bias resulting from insufficient spoiling can result in inaccurate T1 estimates of up to 30% relative to inversion recovery estimated values Stikov et al., 2015. VFA T1 map accuracy and precision is also strongly dependent on the quality of the measured B1 map Lee et al., 2017, which can vary substantially between implementations Boudreau et al., 2017. Modern rapid B1 mapping pulse sequences are not as widely available as VFA, resulting in some groups attempting alternative ways of removing the bias from the T1 maps like generating an artificial B1 map through the use of image processing techniques Liberman et al., 2013 or omitting B1 correction altogether Yuan et al., 2012. The latter is not recommended, because most MRI scanners have default pulse sequences that, with careful protocol settings, can provide B1 maps of sufficient quality very rapidly Boudreau et al., 2017Samson et al., 2006Wang et al., 2005.

Despite some drawbacks, VFA is still one of the most widely used T1 mapping methods in research. Its rapid acquisition time, rapid image processing time, and widespread availability makes it a great candidate for use in other quantitative imaging acquisition protocols like quantitative magnetization transfer imaging Cercignani et al., 2005Yarnykh, 2002 and dynamic contrast enhanced imaging Li et al., 2018Sung et al., 2013.

Keywords:MRIqMRIQuantitativeT1Relaxometry

2About

This Jupyter Book is a series of interactive tutorials about quantitative T1 mapping, powered by qMRLab Karakuzu et al., 2020.

3Inversion Recovery T1 Mapping

Widely considered the gold standard for T1 mapping, the inversion recovery technique estimates T1 values by fitting the signal recovery curve acquired at different delays after an inversion pulse (180°). In a typical inversion recovery experiment (Figure 1), the magnetization at thermal equilibrium is inverted using a 180° RF pulse. After the longitudinal magnetization recovers through spin-lattice relaxation for predetermined delay (“inversion time”, TI), a 90° excitation pulse is applied, followed by a readout imaging sequence (typically a spin-echo or gradient-echo readout) to create a snapshot of the longitudinal magnetization state at that TI.

Inversion recovery was first developed for NMR in the 1940s Hahn, 1949Drain, 1949, and the first T1 map was acquired using a saturation-recovery technique (90° as a preparation pulse instead of 180°) by Pykett & Mansfield, 1978. Some distinct advantages of inversion recovery are its large dynamic range of signal change and an insensitivity to pulse sequence parameter imperfections Stikov et al., 2015. Despite its proven robustness at measuring T1, inversion recovery is scarcely used in practice, because conventional implementations require repetition times (TRs) on the order of 2 to 5 T1 Steen et al., 1994, making it challenging to acquire whole-organ T1 maps in a clinically feasible time. Nonetheless, it is continuously used as a reference measurement during the development of new techniques, or when comparing different T1 mapping techniques, and several variations of the inversion recovery technique have been developed, making it practical for some applications Messroghli et al., 2004Piechnik et al., 2010.

Figure 1. Pulse sequence of an inversion recovery experiment.


4Signal Modelling

The steady-state longitudinal magnetization of an inversion recovery experiment can be derived from the Bloch equations for the pulse sequence {θ180 – TI – θ90 – (TR-TI)}, and is given by:

Mz(TI)=M01−cos(θ180)e−TRT1−[1−cos(θ180)]e−TIT11−cos(θ180)cos(θ90)e−TRT1M_{z}(TI) = M_0 \frac{1-\text{cos}(\theta_{180})e^{- \frac{TR}{T_1}} -[1-\text{cos}(\theta_{180})]e^{- \frac{TI}{T_1}}}{1 - \text{cos}(\theta_{180}) \text{cos}(\theta_{90}) e^{- \frac{TR}{T_1}}}

where Mz is the longitudinal magnetization prior to the θ90 pulse. If the in-phase real signal is desired, it can be calculated by multiplying Eq. 1 by ksin(θ90)e-TE/T 2, where k is a constant. This general equation can be simplified by grouping together the constants for each measurements regardless of their values (i.e. at each TI, same TE and θ90 are used) and assuming an ideal inversion pulse:

Mz(TI)=C(1−2e−TIT1+e−TRT1)M_z(TI) = C(1-2e^{- \frac{TI}{T_1}} + e^{- \frac{TR}{T_1}})

where the first three terms and the denominator of Eq. 1 have been grouped together into the constant C. If the experiment is designed such that TR is long enough to allow for full relaxation of the magnetization (TR > 5T1), we can do an additional approximation by dropping the last term in Eq. 2:

Mz(TI)=C(1−2e−TIT1)M_z(TI) = C(1-2e^{- \frac{TI}{T_1}})

The simplicity of the signal model described by Eq. 3, both in its equation and experimental implementation, has made it the most widely used equation to describe the signal evolution in an inversion recovery T1 mapping experiment. The magnetization curves are plotted in Figure 2 for approximate T1 values of three different tissues in the brain. Note that in many practical implementations, magnitude-only images are acquired, so the signal measured would be proportional to the absolute value of Eq. 3.

Loading...
Loading...

Figure 1:Inversion recovery curves (Eq. 2) for three different T1 values, approximating the main types of tissue in the brain.

Practically, Eq. 1 is the better choice for simulating the signal of an inversion recovery experiment, as the TRs are often chosen to be greater than 5T1 of the tissue-of-interest, which rarely coincides with the longest T1 present (e.g. TR may be sufficiently long for white matter, but not for CSF which could also be present in the volume). Equation 3 also assumes ideal inversion pulses, which is rarely the case due to slice profile effects. Figure 3 displays the inversion recovery signal magnitude (complete relaxation normalized to 1) of an experiment with TR = 5 s and T1 values ranging between 250 ms to 5 s, calculated using both equations.

Loading...
Loading...

Figure 2:Signal recovery curves simulated using Eq. 3 (solid) and Eq. 1 (dotted) with a TR = 5 s for T1 values ranging between 0.25 to 5 s.

5Data Fitting

Several factors impact the choice of the inversion recovery fitting algorithm. If only magnitude images are available, then a polarity-inversion is often implemented to restore the non-exponential magnitude curves (Figure 3) into the exponential form (Figure 2). This process is sensitive to noise due to the Rician noise creating a non-zero level at the signal null. If phase data is also available, then a phase term must be added to the fitting equation Barral et al., 2010. Equation 3 must only be used to fit data for the long TR regime (TR > 5T1), which in practice is rarely satisfied for all tissues in subjects.

Early implementations of inversion recovery fitting algorithms were designed around the computational power available at the time. These included the “null method” Pykett et al., 1983, assuming that each T1 value has unique zero-crossings (see Figure 2), and linear fitting of a rearranged version of Eq. 3 on a semi-log plot Fukushima, 1981. Nowadays, a non-linear least-squares fitting algorithm (e.g. Levenberg-Marquardt) is more appropriate, and can be applied to either approximate or general forms of the signal model (Eq. 3 or Eq. 1). More recent work Barral et al., 2010 demonstrated that T1 maps can also be fitted much faster (up to 75 times compared to Levenberg-Marquardt) to fit Eq. 1 – without a precision penalty – by using a reduced-dimension non-linear least squares (RD-NLS) algorithm. It was demonstrated that the following simplified 5-parameter equation can be sufficient for accurate T1 mapping:

S(TI)=a+be−TIT1S(TI) = a + be^{- \frac{TI}{T_1}}

where a and b are complex values. If magnitude-only data is available, a 3-parameter model can be sufficient by taking the absolute value of Eq. 4. While the RD-NLS algorithms are too complex to be presented here (the reader is referred to the paper, (Barral et al. 2010)), the code for these algorithms was released open-source along with the original publication, and is also available as a qMRLab T1 mapping model. One important thing to note about Eq. 4 is that it is general – no assumption is made about TR – and is thus as robust as Eq. 1 as long as all pulse sequence parameters other than TI are kept constant between each measurement. Figure 4 compares simulated data (Eq. 1) using a range of TRs (1.5T1 to 5T1) fitted using either RD-NLS & Eq. 4 or a Levenberg-Marquardt fit of Eq. 2.

Loading...
Loading...

Figure 3:Fitting comparison of simulated data (blue markers) with T1 = 1 s and TR = 1.5 to 5 s, using fitted using RD-NLS & Eq. 4 (green) and Levenberg-Marquardt & Eq. 2 (orange, long TR approximation).

Figure 5 displays an example brain dataset from an inversion recovery experiment, along with the T1 map fitted using the RD-NLS technique.

Loading...
Loading...

Figure 4:Example inversion recovery dataset of a healthy adult brain (left). Inversion times used to acquire this magnitude image dataset were 30 ms, 530 ms, 1030 ms, and 1530 ms, and the TR used was 1550 ms. The T1 map (right) was fitted using a RD-NLS algorithm.

8Variable Flip Angle T1 Mapping

Variable flip angle (VFA) T1 mapping Christensen et al., 1974Fram et al., 1987Gupta, 1977, also known as Driven Equilibrium Single Pulse Observation of T1 (DESPOT1) Homer & Beevers, 1985Deoni et al., 2003, is a rapid quantitative T1 measurement technique that is widely used to acquire 3D T1 maps (e.g. whole-brain) in a clinically feasible time. VFA estimates T1 values by acquiring multiple spoiled gradient echo acquisitions, each with different excitation flip angles (θ n for n = 1, 2, .., N and θ i ≠ θ j). The steady-state signal of this pulse sequence (Figure 1) uses very short TRs (on the order of magnitude of 10 ms) and is very sensitive to T1 for a wide range of flip angles.

VFA is a technique that originates from the NMR field, and was adopted because of its time efficiency and the ability to acquire accurate T1 values simultaneously for a wide range of values Christensen et al., 1974Gupta, 1977. For imaging applications, VFA also benefits from an increase in SNR because it can be acquired using a 3D acquisition instead of multislice, which also helps to reduce slice profile effects. One important drawback of VFA for T1 mapping is that the signal is very sensitive to inaccuracies in the flip angle value, thus impacting the T1 estimates. In practice, the nominal flip angle (i.e. the value set at the scanner) is different than the actual flip angle experienced by the spins (e.g. at 3.0 T, variations of up to ±30%), an issue that increases with field strength. VFA typically requires the acquisition of another quantitative map, the transmit RF amplitude (B1+, or B1 for short), to calibrate the nominal flip angle to its actual value because of B1 inhomogeneities that occur in most loaded MRI coils Sled & Pike, 1998. The need to acquire an additional B1 map reduces the time savings offered by VFA over saturation-recovery techniques, and inaccuracies/imprecisions of the B1 map are also propagated into the VFA T1 map Boudreau et al., 2017Lee et al., 2017.

Figure 1. Simplified pulse sequence diagram of a variable flip angle (VFA) pulse sequence with a gradient echo readout. TR: repetition time, θn: excitation flip angle for the nth measurement, IMG: image acquisition (k-space readout), SPOIL: spoiler gradient.


9Signal Modelling

The steady-state longitudinal magnetization of an ideal variable flip angle experiment can be analytically solved from the Bloch equations for the spoiled gradient echo pulse sequence {θ n–TR}:

Mz(θn)=M01−e−TRT11−cos(θn)e−TRT1sin(θn)(1)\tag{1} M_{z}(\theta_n) = M_0 \frac{1-e^{- \frac{TR}{T_1}}}{1-\text{cos}(\theta_n) e^{- \frac{TR}{T_1}}} \text{sin}(\theta_n)

where M z is the longitudinal magnetization,

M

0 is the magnetization at thermal equilibrium, TR is the pulse sequence repetition time (Figure 1), and θ n is the excitation flip angle. The M z curves of different T1 values for a range of θ n and TR values are shown in Figure 2.

Loading...
Loading...

Figure 6:Variable flip angle technique signal curves (Eq. 1) for three different T1 values, approximating the main types of tissue in the brain at 3T.

From Figure 2, it is clearly seen that the flip angle at which the steady-state signal is maximized is dependent on the T1 and TR values. This flip angle is a well known quantity, called the Ernst angle Ernst & Anderson, 1966, which can be solved analytically from Equation 1 using properties of calculus:

θErnst=acos(e−TRT1)(2)\tag{2} \theta_{Ernst} = \text{acos}(e^{- \frac{TR}{T_1}})

The closed-form solution (Equation 1) makes several assumptions which in practice may not always hold true if care is not taken. Mainly, it is assumed that the longitudinal magnetization has reached a steady state after a large number of TRs, and that the transverse magnetization is perfectly spoiled at the end of each TR. Bloch simulations – a numerical approach at solving the Bloch equations for a set of spins at each time point – provide a more realistic estimate of the signal if the number of repetition times is small (i.e. a steady-state is not achieved). As can be seen from Figure 3, the number of repetitions required to reach a steady state not only depends on T1, but also on the flip angle; flip angles near the Ernst angle need more TRs to reach a steady state. Preparation pulses or an outward-in k-space acquisition pattern are typically sufficient to reach a steady state by the time that the center of k-space is acquired, which is where most of the image contrast resides.

Loading...
Loading...

Figure 7:Signal curves simulated using Bloch simulations (orange) for a number of repetitions ranging from 1 to 150, plotted against the ideal case (Equation 1 – blue). Simulation details: TR = 25 ms, T1 = 900 ms, 100 spins. Ideal spoiling was used for this set of Bloch simulations (transverse magnetization was set to 0 at the end of each TR).

Sufficient spoiling is likely the most challenging parameter to control for in a VFA experiment. A combination of both gradient spoiling and RF phase spoiling Bernstein et al., 2004Zur et al., 1991 are typically recommended (Figure 4). It has also been shown that the use of very strong gradients, introduces diffusion effects (not considered in Figure 4), further improving the spoiling efficacy in the VFA pulse sequence Yarnykh, 2010.

Loading...
Loading...

Figure 8:Signal curves estimated using Bloch simulations for three categories of signal spoiling: (1) ideal spoiling (blue), gradient & RF Spoiling (orange), and no spoiling (green). Simulations details: TR = 25 ms, T1 = 900 ms, Te = 100 ms, TE = 5 ms, 100 spins. For the ideal spoiling case, the transverse magnetization is set to zero at the end of each TR. For the gradient & RF spoiling case, each spin is rotated by different increments of phase (2𝜋 / # of spins) to simulate complete decoherence from gradient spoiling, and the RF phase of the excitation pulse is ɸn = ɸn-1 + nɸ0 = ½ ɸ0(n2 + n + 2) Bernstein et al., 2004 with ɸ0 = 117° Zur et al., 1991 after each TR.

Acknowledgments

This work was supported by the Canadian Open Neuroscience Platform (CONP), the Quebec Bio-Imaging Network (QBIN), and the Montreal Heart Institute Foundation (MHI).

References
  1. Boudreau, M., Keenan, K. E., & Stikov, N. (2020). Chapter 2 - Quantitative T1 and T1ρ Mapping. In N. Seiberlich, V. Gulani, F. Calamante, A. Campbell-Washburn, M. Doneva, H. H. Hu, & S. Sourbron (Eds.), Quantitative Magnetic Resonance Imaging (Vol. 1, pp. 19–45). Academic Press. 10.1016/B978-0-12-817057-1.00004-4
  2. Boudreau, M., Keenan, K. E., & Stikov, N. (2023). Quantitative T1 MRI. 10.55458/neurolibre.00019
  3. Karakuzu, A., DuPre, E., Tetrel, L., Bermudez, P., Boudreau, M., Chin, M., Poline, J.-B., Das, S., Bellec, P., & Stikov, N. (2022). NeuroLibre : A preprint server for full-fledged reproducible neuroscience. OSF Preprints. 10.31219/osf.io/h89js
  4. DuPre, E., Holdgraf, C., Karakuzu, A., Tetrel, L., Bellec, P., Stikov, N., & Poline, J.-B. (2022). Beyond advertising: New infrastructures for publishing integrated research objects. PLOS Computational Biology, 18(1), e1009651. 10.1371/journal.pcbi.1009651
  5. Harding, R. J., Bermudez, P., Bernier, A., Beauvais, M., Bellec, P., Hill, S., Karakuzu, A., Knoppers, B. M., Pavlidis, P., Poline, J.-B., Roskams, J., Stikov, N., Stone, J., Strother, S., Consortium, C., & Evans, A. C. (2023). The Canadian Open Neuroscience Platform—An open science framework for the neuroscience community. PLOS Computational Biology, 19(7), 1–14. 10.1371/journal.pcbi.1011230
  6. Karakuzu, A., Boudreau, M., Duval, T., Boshkovski, T., Leppert, I., Cabana, J.-F. ̧cois, Gagnon, I., Beliveau, P., Pike, G., Cohen-Adad, J., & Stikov, N. (2020). qMRLab: Quantitative MRI analysis, under one umbrella. J. Open Source Softw., 5(53), 2343. 10.21105/joss.02343
  7. Hahn, E. L. (1949). An accurate nuclear magnetic resonance method for measuring spin-lattice relaxation times. Physical Review, 76(1), 145. 10.1103/PhysRev.76.145
  8. Drain, L. (1949). A direct method of measuring nuclear spin-lattice relaxation times. Proceedings of the Physical Society. Section A, 62(5), 301. 10.1088/0370-1298/62/5/306
  9. Pykett, I., & Mansfield, P. (1978). A line scan image study of a tumorous rat leg by NMR. Physics in Medicine & Biology, 23(5), 961. 10.1088/0031-9155/23/5/012
  10. Stikov, N., Boudreau, M., Levesque, I. R., Tardif, C. L., Barral, J. K., & Pike, G. B. (2015). On the accuracy of T1 mapping: Searching for common ground. Magnetic Resonance in Medicine, 73(2), 514–522. 10.1002/mrm.25135
  11. Steen, R. G., Gronemeyer, S. A., Kingsley, P. B., Reddick, W. E., Langston, J. S., & Taylor, J. S. (1994). Precise and accurate measurement of proton T1 in human brain in vivo: validation and preliminary clinical application. Journal of Magnetic Resonance Imaging, 4(5), 681–691. 10.1002/jmri.1880040511
  12. Messroghli, D. R., Radjenovic, A., Kozerke, S., Higgins, D. M., Sivananthan, M. U., & Ridgway, J. P. (2004). Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magnetic Resonance in Medicine: An Official Journal of the International Society for Magnetic Resonance in Medicine, 52(1), 141–146. 10.1002/mrm.20110
  13. Piechnik, S. K., Ferreira, V. M., Dall’Armellina, E., Cochlin, L. E., Greiser, A., Neubauer, S., & Robson, M. D. (2010). Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. Journal of Cardiovascular Magnetic Resonance, 12(1), 1–11. 10.1186/1532-429X-12-69
  14. Christensen, K. A., Grant, D. M., Schulman, E. M., & Walling, C. (1974). Optimal determination of relaxation times of Fourier transform nuclear magnetic resonance. Determination of spin-lattice relaxation times in chemically polarized species. The Journal of Physical Chemistry, 78(19), 1971–1977. 10.1021/j100612a022
  15. Fram, E. K., Herfkens, R. J., Johnson, G. A., Glover, G. H., Karis, J. P., Shimakawa, A., Perkins, T. G., & Pelc, N. J. (1987). Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magnetic Resonance Imaging, 5(3), 201–208. 10.1016/0730-725x(87)90021-x